Operation
English
Figure A3 - DPI 610/615 IS HC - Flushing/Filling Connections
(b)
Fill the priming fluid container with fresh hydraulic fluid of the required
type. Use the priming pump, to pump fresh hydraulic fluid through the
system until clean hydraulic fluid, free of air bubbles, emerges into the
container at the output port. Discard the contaminated fluid expressed
during this process.
(c)
Remove the hose located to the output port and, to prevent the ingress of
any contaminant, fit a blanking plug in its place.
(d)
Close the isolation valve and remove the priming fluid inlet hose from the
priming inlet.
(e)
Clean off any surplus oil which may have spilled onto the instrument casing.
The instrument is now ready for operation or storage. If storing, apply a label detailing
the type of hydraulic fluid contained in the instrument.
For long term storage, it is
recommended that the instrument be drained and stored empty.
Flushing - Replenishing or Changing the Hydraulic Fluid
If necessary, to remove any contaminants, flush out the hydraulic system as
follows.
(a)
Connect a priming hose assembly to the fluid inlet port and the pressure
port as shown below.
Summary of Contents for DPI 610 IS
Page 4: ...Page 4 of 68 K249 Issue No 4 Do Not Print This Page ...
Page 5: ...Druck DPI 610 615 IS Portable Pressure Calibrator User manual K249 GE Infrastructure Sensing ...
Page 6: ......
Page 9: ...iii K249 Issue No 4 English English ...
Page 57: ...English 43 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION ...
Page 58: ...44 K249 Issue No 4 This page is intentionally blank ...
Page 64: ...50 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION Operation Notes ...
Page 65: ...English 51 K249 Issue No 4 Appendix 1 DATALOG FILE EXAMPLE ...
Page 66: ...52 K249 Issue No 4 This page intentionally blank ...