42
K249 Issue No. 4
Calibration
Temperature
•
Precision Temperature Meter.
On completion of calibration routines, exit the calibration mode by pressing the
EXIT
function key.
Change PIN Number
To change the PIN number, select
CHANGE PIN
from the calibration menu and
proceed as follows. Note that to set and verify a new PIN number, the new PIN
number must be entered twice.
If the second number entered differs from the first, no new PIN number is set.
Add External Sensor
To add an external sensor to the directory of external sensors, proceed as follows.
•
Plug the external transducer into the connector located on top of
the instrument and proceed as shown below.
Digitally-compensated transducers will be downloaded into the directory as soon
as the F1 key is operated. Data for other types can be edited by selecting EDIT
(F2). If the data cannot be edited a warning message (ACTION INVALID) is
displayed.
Cleaning Instructions
Clean the case of the instrument by using a damp cloth and mild detergent.
Note:
DO NOT
use any solvents to clean the instrument.
EXT SENSORS
SERIAL No.
RANGE
bar
TYPE
ADD NEW
SENSOR
EXT SENSORS
SERIAL No.
RANGE
bar
TYPE
1154738
20.000
g
ADD NEW
SENSOR
EDIT
EXT SENSORS
SERIAL No.
RANGE
bar
TYPE
1154738
20.000
g
REMOVE
CHANGE
DETAILS
Summary of Contents for DPI 610 IS
Page 4: ...Page 4 of 68 K249 Issue No 4 Do Not Print This Page ...
Page 5: ...Druck DPI 610 615 IS Portable Pressure Calibrator User manual K249 GE Infrastructure Sensing ...
Page 6: ......
Page 9: ...iii K249 Issue No 4 English English ...
Page 57: ...English 43 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION ...
Page 58: ...44 K249 Issue No 4 This page is intentionally blank ...
Page 64: ...50 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION Operation Notes ...
Page 65: ...English 51 K249 Issue No 4 Appendix 1 DATALOG FILE EXAMPLE ...
Page 66: ...52 K249 Issue No 4 This page intentionally blank ...