K249 Issue No. 4
24
Select Output
To select an output channel for display, select
ADVANCED
mode from the Task
menu and proceed as follows. If a channel has a range of units available, a
UNITS
soft box (actioned by the F2 function key), will be also be written to the display.
The following procedure shows the method of output channel selection. Note that
there are two pages of options. The second page can be obtained directly from
the first by pressing the
RIGHT
cursor key.
Note:
Left/right arrow keys function as page up/down keys.
To change the output units (
Pressure
channels only), select the channel with the
cursor keys and press F2 before pressing
ENTER
.
Electrical Outputs (Loop Power)
All the electrical outputs, the output loop must be powered from an external
supply (current sinking).
External connections to the front panel of the instrument are shown below.
Advanced Task
Select Output
55mA
Vin
mA in
mA Sink
max 30V
CAT II
CURRENT SINKING
LOAD
External
d.c. supply
(30V max)
max
+
-
BARRIER
SAFE
SAFE
Summary of Contents for DPI 610 IS
Page 4: ...Page 4 of 68 K249 Issue No 4 Do Not Print This Page ...
Page 5: ...Druck DPI 610 615 IS Portable Pressure Calibrator User manual K249 GE Infrastructure Sensing ...
Page 6: ......
Page 9: ...iii K249 Issue No 4 English English ...
Page 57: ...English 43 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION ...
Page 58: ...44 K249 Issue No 4 This page is intentionally blank ...
Page 64: ...50 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION Operation Notes ...
Page 65: ...English 51 K249 Issue No 4 Appendix 1 DATALOG FILE EXAMPLE ...
Page 66: ...52 K249 Issue No 4 This page intentionally blank ...