17
K249 Issue No. 4
English
On completion of
Tx SIM
set-up, the display is configured as follows (e.g.).
Connect an external power source to the output loop as detailed on Page 24.
To subsequently change any of the
Tx SIM
scaling parameters, press
CHANGE
VALUE
key (F1) to obtain the TX Simulation set-up display.
To change the pressure units, press
INPUT
and select the required scale units. If
the required scale units are not listed, press
SETUP
, select
SETTINGS
and proceed
as detailed on Page 37.
Taking Measurements
Transmitter Simulator (TX SIM) Task
When used with an external voltage source (see Page 24), provides a current
output proportional to the calibrator’s measured output pressure (indicated
pressure on indicator only version). Select task
TX SIM.
Press
EXIT
to skip set-up
screen if parameters are correct.