v
K249 Issue No. 4
English
Specification
Pressure Ranges (Internal Transducers)
Safe working pressure
20 bar range
1.75 x full-scale
350 bar range
1.2 x full-scale
400 bar range
1.5 x full-scale
All other ranges
2 x full-scale
Accuracy
Combined non-linearity, hysteresis and repeatability
±70 mbar range
0.05% F.S.
up to ±150 mbar
0.05% span
200 mbar to 20 bar (Calibrator):
0.025% F.S.
35 bar to 700 bar (Indicator):
0.025% F.S.
70 bar to 400 bar (Hydraulic)
0.025% F.S.
Temperature Effects
±0.004% of reading/°C (averaged over -10° to -40°C w.r.t. 20°C)
Electrical Parameters
Voltage Inputs
Range:
±30V
Accuracy
±0.05% rdg, ±0.004% F.S.
Resolution
100µV max
Current Inputs
Range:
±55mA
Accuracy
±0.05% rdg, ±0.004% F.S.
Resolution
1µA max
Current sink
Range:
24mA
Accuracy
±0.05% rdg, ±0.01% F.S.
Display
Size:
60 x 60 mm LCD Graphics
Reading
±99999, update rate 2 readings/sec
Environmental
Operating Temp.:
-10°C to 50°C (Calibrated -10°C to 40°C)
Storage Temp:
-20°C to 60°C
Calibration Temp:
21°C ±2°C
Sealing
Sealed to IP54
Physical
Size:
300 x 170 x 140 mm
Weight:
3 kg
Summary of Contents for DPI 610 IS
Page 4: ...Page 4 of 68 K249 Issue No 4 Do Not Print This Page ...
Page 5: ...Druck DPI 610 615 IS Portable Pressure Calibrator User manual K249 GE Infrastructure Sensing ...
Page 6: ......
Page 9: ...iii K249 Issue No 4 English English ...
Page 57: ...English 43 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION ...
Page 58: ...44 K249 Issue No 4 This page is intentionally blank ...
Page 64: ...50 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION Operation Notes ...
Page 65: ...English 51 K249 Issue No 4 Appendix 1 DATALOG FILE EXAMPLE ...
Page 66: ...52 K249 Issue No 4 This page intentionally blank ...