K249 Issue No. 4
Appendix 1
Typical Uploaded Datalog File (DPI 610 IS)
The following gives details of a typical data file upload.
FILENAME: TEST 5
DATE:
05/11/1999
TIME:
15:58
TRIGGER:
KEYPRESS
AMBIENT TEMP : 24.1 C
NO. OF POINTS 11
RECORD TYPE 0
CURRENT mA PRESSURE INT bar
3.902 0.008 -0.65 %span 05/11/1999 15:58:55
6.076 2.311 1.42 %span 05/11/1999 16:00:03
7.598 4.404 0.47 %span 05/11/1999 16:00:57
8.085 5.023 0.41 %span 05/11/1999 16:01:45
9.949 7.249 0.94 %span 05/11/1999 16:02:42
13.002 11.300 -0.23 %span 05/11/1999 16:03:30
17.005 16.102 0.77 %span 05/11/1999 16:05:00
17.766 17.106 0.51 %span 05/11/1999 16:06:07
17.812 17.105 0.80 %span 05/11/1999 16:07:49
18.532 17.965 1.00 %span 05/11/1999 16:16:35
20.007 20.212 -1.02 %span 05/11/1999 16:18:23
The
Record Type
number contained in the header is a 16-bit field, with the
individual bits representing result formats as detailed below.
Record Type 1
P-Display
Record Type 2
Switch Test
Record Type 4
Leak Test
Record Type 8
Input channel: Min/Max ON
Record Type 120
Combination of all input process functions
Record Type 896
Combination of all output process functions
Record Type 1024
mA step mode
Record Type 2048
mA value mode
Record Type 4096
mA ramp mode
Record Type 8192
no input selected
Record Type 16384
no output selected
Summary of Contents for DPI 610 IS
Page 4: ...Page 4 of 68 K249 Issue No 4 Do Not Print This Page ...
Page 5: ...Druck DPI 610 615 IS Portable Pressure Calibrator User manual K249 GE Infrastructure Sensing ...
Page 6: ......
Page 9: ...iii K249 Issue No 4 English English ...
Page 57: ...English 43 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION ...
Page 58: ...44 K249 Issue No 4 This page is intentionally blank ...
Page 64: ...50 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION Operation Notes ...
Page 65: ...English 51 K249 Issue No 4 Appendix 1 DATALOG FILE EXAMPLE ...
Page 66: ...52 K249 Issue No 4 This page intentionally blank ...