Typical Calibration Setup (Pressure to Voltage)
Connect a device under test to the instrument as shown below.
•
General Procedure
Use the hand-pump to pressurise the system to the required level as
indicated on the display. Allow the display to settle and screw the volume
adjuster in or out as a fine adjustment to the required pressure. Record the
measurand (e.g.)
Voltage
, reading at each applied pressure.
Zero Display Reading
Both the input and output windows can be set to zero by operation of the
ZERO
key, providing that the displayed reading is already within 5% of zero. To zero
either the
INPUT
(Measurand) or
OUTPUT
(Source) windows, proceed as follows
(e.g.),
Getting Started