20
K249 Issue No. 4
Process Functions
If required, the following process functions are available on the Measurand
(INPUT)
display but
only
in
ADVANCED
task. If the instrument is in any other mode i.e.
BASIC or any other task mode, the input and output displays must first be
configured in
ADVANCED
task.
Note:
PROCESS functions are not available to the output (SOURCE) channel.
A summary of the process functions follows.
•
Tare
Allows either the current display value or a manually
entered value to be tared off display parameter
reading.
•
Max/Min
Displays running Max/Min and Present display values
simultaneously. Resettable via F1 key.
•
Filter
Applies low pass filter function to displayed
parameter. Filter characteristics
(Settling time and
Band)
are user programmable.
•
Flow
Applies square root function to displayed parameter.
•
% Span
Converts displayed parameter reading to a
percentage of span. Span definable via the F1 key.
Following selection of
ADVANCED
from the task menu, press the
INPUT
key. Use
the
Up
or
Down
cursor keys to select the required input. Press the
PROCESS
(F1) key and use the
Left
or
Right
cursor keys to enable the process on/off
(e.g.),
Press
ENTER
to switch the process ON with existing settings or F1 to change
process settings (where applicable).
Advanced Task
Process Functions
Summary of Contents for DPI 610 IS
Page 4: ...Page 4 of 68 K249 Issue No 4 Do Not Print This Page ...
Page 5: ...Druck DPI 610 615 IS Portable Pressure Calibrator User manual K249 GE Infrastructure Sensing ...
Page 6: ......
Page 9: ...iii K249 Issue No 4 English English ...
Page 57: ...English 43 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION ...
Page 58: ...44 K249 Issue No 4 This page is intentionally blank ...
Page 64: ...50 K249 Issue No 4 HYDRAULIC CALIBRATOR VERSION Operation Notes ...
Page 65: ...English 51 K249 Issue No 4 Appendix 1 DATALOG FILE EXAMPLE ...
Page 66: ...52 K249 Issue No 4 This page intentionally blank ...