![GE AxiChrom 50 Operating Instructions Manual Download Page 69](http://html1.mh-extra.com/html/ge/axichrom-50/axichrom-50_operating-instructions-manual_4090700069.webp)
5
Click on Next to display the Slurry and Packing Test Parameters wizard page. Make
the appropriate selections.
Figure 5.4: Intelligent Packing Wizard (4 of 5), ÄKTApilot
Enter the slurry concentration (see Preparing the
slurry, on page 63).
Slurry concentration
Enter an Equilibration volume (measured in CV)
within the designated range. Default values are
given according to the recommended packing and
test liquid for the selected medium.
Equilibration volume
Leave checked to use the suggested default liquid
velocity of 30 cm/h.
Use default liquid velocity
during equilibration
Leave checked to use the suggested default liquid
velocity of 30 cm/h.
Use default liquid velocity
during packing test
Select whether to perform packing tests upflow,
downflow or both. Both are selected by default.
Packing test upflow and
Packing test downflow
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
69
5 Operation
5.1 Intelligent Packing of the column with ÄKTAexplorer and ÄKTApilot