![GE AxiChrom 50 Operating Instructions Manual Download Page 34](http://html1.mh-extra.com/html/ge/axichrom-50/axichrom-50_operating-instructions-manual_4090700034.webp)
Plastic bed support dependent column parts:
Applies to all column sizes unless otherwise stated.
Item
Pos no
1
Adapter plate, plastic bed support
501
Bottom plate, plastic bed support
502
Bed support, 10 µm/20 μm plastic
301, 302
Screw, plastic bed support (50/70)
401
Distributor top, plastic bed support
402
Distributor bottom, plastic bed support
404
Washer, plastic bed support
412
Snap ring top/bottom bed support
421 a,b
Support net, top/bottom bed support (50/70)
422 a,b
Adapter tube id. 1.7, plastic bed support (50)
Adapter tube id. 2.9, plastic bed support (70/100)
405
O-ring, plastic bed support (50/70)
409
O-ring, plastic bed support (50/70)
413
1
Position number in the exploded diagrams in Appendices A-C.
4.4
Wetting instructions for plastic bed supports
Follow these instructions to wet top and bottom plastic bed supports before packing
the column.
1
Before using new plastic bed supports, wet the bed supports in 20% v/v 1-propanol
for at least 12 hours in an appropriate container. Make sure that the bed supports
are submerged below the surface of the solution by placing a weight on them.
2
Rinse the bed supports in 20% ethanol and keep them soaked in 20% ethanol until
they are mounted in the column.
3
Place the wetted bottom bed support in a new snap ring. Fit the bed support and
snap ring assembly onto the bottom distributor and mount on to the column as de-
scribed in this Operating Instruction.
34
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
4 Getting started
4.3 Bed supports