![GE AxiChrom 50 Operating Instructions Manual Download Page 22](http://html1.mh-extra.com/html/ge/axichrom-50/axichrom-50_operating-instructions-manual_4090700022.webp)
3.2
Primary specifications
Table 3.1: Primary column specifications and characteristics.
100 column
70 column
50 column
100
70
50
Inner diameter (mm)
10-30
30-50
5-35
30-60
Bed heights (cm)
Short column tube: Intelligent packing
Long column tube: Intelligent packing
Short column tube: Adapter stroke length
Long column tube: Adapter stroke length
0.79-2.36
2.36-3.93
0.39-1.15
1.15-1.92
0.20-0.59
0.59-0.98
Bed volume (l)
Short column tube
Long column tube
78.5
38.5
19.6
Column cross section (cm
2
)
10
15
20
Max packing pressure (bar g)
1
8
8
10
Max operating pressure (bar g)
2
2-30
Operating temperature (°C)
1-14
Operating pH range
10, 20
Stainless steel bed support pore sizes (µm)
10, 20
Plastic bed support pore sizes (µm)
16.5
19.5
10.0
11.5
6.5
7.5
Column weight, empty (kg)
3
Short column tube
Long column tube
N/A
7.0
7.5
1.0
7.0
7.5
Foot/pivot stand weight (kg)
AxiChrom foot
Pivot stand:
Short column tube
Long column tube
1.65
1.8
1.4
1.7
Max. working height (m)
4
Short column tube
Long column tube
350 x 360 (pivot stand)
352 x 352 (foot)
350 x 360 (pivot stand)
Footprint (mm x mm)
1
Refers to the maximum allowed pressure for the column, and specifically the maximum pressure
in the hydraulic chamber during the packing procedure.
22
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
3 Specifications and characteristics
3.2 Primary specifications