3
Refit the bed support [301] onto the adapter being careful to align the holes around
the circumference of both parts.
NOTICE
The bed support shall be mounted in the position according to
figure B below before the screws [401] are tightened. Damage
to the bed support can occur if the bed support slides on only
partway, as in figure A below. This implies that the O-ring [603]
{411} has snapped off the groove in the adapter for 50 and 70
columns, or off the groove in the end cell for the 100 columns.
Tightening of the screws [401] will cause the O-ring [603] {411}
to be pushed onto the bed support and damage the bed sup-
port.
Approx. 1 mm
A
B
4
For the 100 column only: Check and replace the O-rings {413} if necessary.
5
Use an Allen key to carefully refit the screws [401] on the adapter. Using a torque
driver, tighten each screw a little at a time crosswise to 0.5 Nm for 50 and 70 columns
and 1 Nm for the 100 column. Do not overtighten. When tightened, there will be a
gap of approximately 1 mm between the bed support and the adapter.
116
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
6 Maintenance and service
6.2 Performing maintenance and service