CARESCAPE Monitor B650
198
2062973-002
12.3.11 Detaching the Base unit
Disassemble first:
a.
12.3.2. Detaching the Interface Board
b.
12.3.3. Detaching the Frame Side Housing Decorations
c.
12.3.4. Detaching the Top Cover
d.
12.3.5. Disassembling the main unit into Rear and Front units
e.
12.4.1. Detaching the Front Unit Assembly
f.
12.4.2. Detaching the LCD Display Unit
6. Remove the four T10 screws mounting the
AC/DC board to the back plate.
[Torque 1.6 Nm]
Hold the AC/DC board to prevent it from
falling.
Reassemble in reverse order.
- Attach the 4 screws that mount the AC/DC board [Torque 1.6 Nm].
- Ensure the cable clamp and the spring are in place.
- Reattach the housing back cover so that the 4 snaps locks.
- Attach the 1 screw that mounts the housing back cover.
- Reconnect the AC/DC cable and the protective earth cable to the AC/DC board.
- Place the DC/DC cable in the cable holder.
Summary of Contents for CARESCAPE Monitor B650
Page 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Page 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Page 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Page 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Page 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Page 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Page 177: ...Troubleshooting 169 2062973 002 ...
Page 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Page 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Page 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Page 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Page 239: ......