CARESCAPE Monitor B650
C - 4 (6)
2062973-002
Configuring the patient simulator:
Refer to the simulator documentation for details on how to use and configure the simulator.
6.
Configure the patient simulator to output ECG waveform with:
−
ECG rhythm: a normal sinus rhythm
−
Heart rate: 80 bpm
−
Amplitude: 1 mV
Configuring the patient monitor:
7.
Configure the ECG1, ECG2 and ECG3 waveform fields to the monitor screen with
adequate priority.
8.
In the
Parameter Setup > ECG > Setup
, select:
−
ECG1 lead: II
−
ECG2 lead: V1
−
ECG3 lead: aVL
.
9.
Start a patient case / admit a patient in
Data & Pages > Admit/Discharge
menu.
10. Select
Data & Pages > Other Patients
. Select the unit and bed name of the stationary
monitor and then select
View
.
Configuring the service PC:
11. Login to Webmin.
12. Select
Diagnostics > WLAN
.
NOTE: The WLAN information shown in Webmin Diagnostics page is static. Press
F5
to refresh
the web browser and to see up-to-date WLAN diagnostics data.
Test execution
Execute the test procedure according to the test plan. Contact the nursing staff to ensure
access to the needed areas before starting the test.
1.
Move the roll cart to the starting point of the planned test route.
2.
Stop at each test point and perform the following tasks:
a.
On the transport monitor:
•
Verify that ECG waveforms from the remote, stationary patient monitor display in
the bed-to-bed view without any losses.
b.
On the service PC: Refresh Webmin view by selecting
F5
and:
•
Identify the Access Point the patient monitor is connected to.
•
Verify that the
Signal level (RSSI)
in dBm is greater than or equal to -60 dBm.
•
Verify that the
Transmit Rate in Mbps
is greater than or equal to 5.5 Mbps.
c.
Mark the network time, RSSI and Transmit Rate to the test form.
d.
If there is a waveform loss situation or the RSSI or the Transmit Rate is lower than
specified:
•
Observe the length of the waveform loss and, if possible, potential cause of it, for
example roaming or out of range situations.
Summary of Contents for CARESCAPE Monitor B650
Page 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Page 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Page 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Page 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Page 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Page 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Page 177: ...Troubleshooting 169 2062973 002 ...
Page 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Page 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Page 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Page 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Page 239: ......