CARESCAPE Monitor B650
150
2062973-002
To download log files:
1.
Log in to Webmin.
2.
Select
Diagnostics > Download Logs
.
3.
Select the log(s) you want to download.
4.
Select
Download
.
5.
According to your Webmin access:
a.
If you are using Webmin on a service PC, you can save the log file to any storage device
connected to the service PC:
•
In the
File Download
dialog box, select
Save
.
•
In the
Save as
dialog box, select the destination drive and folder and select
Save
.
NOTE: You may change the default filename, but do not change the file extension.
b.
If you are using Webmin locally through the integrated browser, you can save the settings
file to a USB flash drive that is connected to one of the patient monitor's USB ports:
•
The
Download Logs
window will show you the name of the created log file.
•
Select
Download
to save the log file to the USB flash drive.
NOTE: Do not disconnect the USB storage device until downloading is complete.
6.
Send this log file to GE Service for further investigation.
To view log files:
1.
Log in to Webmin.
2.
Select
Diagnostics
>
View Logfiles
.
PDM log
All PDM errors and messages.
System log
•
OS events and errors, including operating system related
information, such as clinical application startup and recovery
information, power on self-test results, etc.
•
UIC, PMC, WLAN and battery events and errors.
•
InSite with ExC events and errors, including InSite with ExC
agent related information.
Clinical log
•
Clinical-related application events, including module
(parameter) connections/disconnections, case starts/ends,
cold starts, warm starts, etc.
•
Technical notes and errors displayed for clinical users,
including all host and module related technical errors (e.g.,
‘Failure in Agent ID’).
•
Clinical alarms, including clinical application related patient
alarms and their level (e.g., ‘FiO
2
Low’).
•
Clinical user interactions, such as the host keystrokes and
touchscreen selections, menu setting changes, etc.
•
EPI layer related errors, including information about EPI layer
issues detected by the clinical application.
•
Host serial number and active software versions.
Log file name
Contents
Summary of Contents for CARESCAPE Monitor B650
Page 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Page 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Page 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Page 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Page 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Page 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Page 177: ...Troubleshooting 169 2062973 002 ...
Page 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Page 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Page 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Page 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Page 239: ......