12.3.9 Detaching the CPU Board
Disassemble first:
a.
12.3.2. Detaching the Interface Board
b.
12.3.3. Detaching the Frame Side Housing Decorations
c.
12.3.4. Detaching the Top Cover
d.
12.3.5. Disassembling the main unit into Rear and Front units
e.
12.3.8. Detaching the DC/DC board
1. Detach the WLAN card from the CPU if
installed.
NOTE: Be careful not to damage the antenna
cables.
2. Detach the uDOM.
IMPORTANT: Detach the original uDOM and
attach it to the new CPU Board.
The software, licenses and clinical and service
configurations are stored in the uDOM.
3. Detach the Hinge Flex Board and the Display
Cable from the CPU Board.
NOTE: Be careful not to damage the Hinge Flex
Board.
4. Detach the Sync cable, the Loudspeaker
cable, and the U/I cable from the CPU Board.
5. Remove the nine T10 screws that hold the
CPU Board to the mid-frame.
6. Detach the CPU Board. [Torque 0.8 Nm]
Reassemble in reverse order.
- Insert the WLAN card into the card slot.
- Place the antenna cables carefully inside the cover.
NOTE: Center align the user interface cable carefully to the connector in the CPU Board and/or User Interface
Board.
Note the alignment pins to set the CPU Board in the right place.
NOTE: After replacing the CPU Board, during the first start-up, the patient monitor will automatically check the
EMBC software version in the replaced CPU Board and update the software if necessary. Wait for 5 minutes to
see if the software update is initiated and do not interrupt the process.
UI cable
Loudspeaker
Sync
cable cable
Display
cable
Hinge
cable
Flex
uDOM
Summary of Contents for CARESCAPE Monitor B650
Page 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Page 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Page 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Page 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Page 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Page 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Page 177: ...Troubleshooting 169 2062973 002 ...
Page 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Page 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Page 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Page 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Page 239: ......