NOTE: Make sure that at least one other patient monitor is on the network. The other patient
monitor must be in an admitted state and have an active ECG measurement with a simulator
signal.
1.
Disconnect the network cable from the MC Network connector, if connected.
2.
Check that the wireless network symbol and the signal strength symbol are shown
and that the signal strength is adequate.
3.
Select
Data & Pages – Other Patients
and select a patient from the list.
4.
Select
View
and verify that a window with parameters from another patient displays on
the left side of the screen.
5.
Select
Close View
to close the window.
6.
Reconnect the MC Network cable back to the MC Network connector, if applicable.
8.3.11 IX printers
Perform the following test only if the patient monitor is connected to a printer in the IX Network
and you did not print a test page while you configured the IX printer.
1.
Log in to Webmin.
2.
Select
Configuration
>
Printers > Print Test Page
.
3.
Select the IX Printer to test.
4.
Select
Save
.
5.
Verify that the test page was printed to the selected printer.
6.
Repeat steps 3 to 5 for all connected IX printers.
8.3.12 Insite with EXC
Perform the following test only if the remote service is configured and enabled.
•
Contact your local online support center to find out if they can view the patient monitor.
8.3.13 Test completion
Discharge Patient
or
Reset Case
to discard any changes made to the patient monitor
configuration during checkout.
•
Complete the Appendix
Summary of Contents for CARESCAPE Monitor B650
Page 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Page 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Page 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Page 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Page 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Page 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Page 177: ...Troubleshooting 169 2062973 002 ...
Page 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Page 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Page 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Page 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Page 239: ......