Disassembly and reassembly
187
2062973-002
12.3 Main unit disassembly
12.3.1 Replacing the mains fuses
The mains fuses are situated on the back of the
monitor, below the mains connector.
1. Lever the screwdriver against the cable holder
to release the fuse holder and pull it out of the
frame.
2. Pull the fuses out of the fuse holder and insert
new ones.
3. Push the fuse holder all the way in until it
snaps to its place.