C:V RATIO MENU (COMPRESSION TO
VENTILATION RATIO)
Adjust the compression to ventilation ratio
using the + and - buttons. Press OK to save
the changes and return to previous screen.
The default value for the C:V Ratio is 30
compressions to 2 ventilations.
COMPRESSION / VENTILATION RATE
MENU
The default value for the compression rate is
100 compressions per minute. Adjust rate using
the ‘+’ and ‘-’ buttons. Press OK to save the
changes and return to previous screen. The
compression rate limit is 50 - 150 CPM. The
ventilation rate limit is 2- 60 VPM.
CALIBRATION MENU
Press CALIB. to access additional menu
selections.
CALIBRATING CHEST COMPRESSIONS
The simulator is pre-calibrated to current CPR
guidelines at time of manufacture. If the CPR
guidelines change, calibrate the sensors inside
the simulator using this option.
Calibrating chest compressions:
1. Select C CAL.
2. Press START to begin the calibration
procedure.
3. Omni will ask you to perform several correct
chest compressions. The facilitator should
follow the text cue on the screen to perform
just one compression at a time until inished.
4. When Omni is ready to calibrate it will
display COMPRESS.
5. Perform a correct compression.
6. Follow the text cue on the screen to perform
the remaining four compressions. When the
calibration is complete Omni will display
DONE.
7. Press SAVE.