Troubleshooting
GENERAL TROUBLESHOOTING GUIDE
Use the following table to ind causes and solutions to a number of possible problems.
Symptom
Possible Cause
Solution
Omni is not turning on
Power supply not connected to
simulator
Connect the power supply
cable to the simulator and the
other end to a power source
Communication cable not
connected to Omni or the
simulator
Connect the communication
cable to the simulator and to
Omni.
I lost communication with the
simulator
Corrupted connection
Reboot Omni by unplugging
the communication cable
and reconnecting after a few
seconds
Communication never gets
established or is lost
Data cable is not connected
Ensure the Ethernet cable is
plugged into Omni and the
simulator
Verify the power cable is
connected to the simulator
Omni is not detecting any
compression or ventilations
System is not calibrated
Calibrate compressions and
ventilations using Omni
Omni detects compressions
but no ventilations
Module in the simulator is
disconnected
Contact technical support to
troubleshoot problem
Omni takes too long to boot up A system restore is required
1. Unplug the communication
cable from Omni.
2. Press and hold down the
two outer keys, located on the
right side of the controller, while
reconnecting the communication
cable.
3. Once a “Please Wait” message
appears on the screen, release the
two outer keys.
Wait for the main screen to load
and continue with simulations.