30 |
User Guide | Omni™ Code Blue® and BP Arm |
VOLUME MENU
Increase or lower the volume of Korotkoff
sounds using the
VOLUME
menu. Press
OK
to save the changes and return to previous
screen. The default value for the Korotkoff
sounds volume is 5.
CALIB MENU
Calibrate the sensors inside the simulator
using this option. Set the standard against
which auscultation will be evaluated during the
exercise. Blood pressure cuff calibration should
be performed only when the Korotkoff sounds
do not match the systolic and diastolic values
set from Omni.
It is not necessary to re-calibrate Omni
for each training session. Omni will
remember previous calibration settings.
CALIBRATION INSTRUCTIONS
1. Assemble the system and turn Omni on.
2. Place the blood pressure cuff on the arm.
3. Connect the cuff’s extension to connector on
the control panel.
4.
Select the
CALIB
menu from the main
screen.
5. Press NEXT to begin the calibration
procedure.
6. You will be prompted to set the BP cuff
to a certain pressure. Hold that pressure
constant and press the NEXT button.
7. Omni will now prompt you to set the BP cuff
to 20 mmHg. Hold that pressure constant
and press the NEXT button.
Follow the text cue on the screen and repeat the
procedure for each pressure level. The facilitator
should follow the text cue on the screen to inlate
the cuff just one value at a time, until inished.
After calibrating the sensor, Omni will return to
the main screen and display the reading shown
on the sphygmomanometer.