28 |
User Guide | Omni™ Code Blue® and BP Arm |
CALIBRATING AIRWAY VENTILATIONS
Set the standard against which ventilation will
be evaluated during the exercise.
Calibrating airway ventilations:
1. Select V CAL.
2. Press START to begin the calibration
procedure.
3. Omni will ask you to perform a number of
correct airway ventilations. The facilitator
should follow the text cue on the screen to
perform just one ventilation at a time, until
inished.
4. When Omni is ready to calibrate it will
display VENTILATE.
5. Perform a ventilation.
6. Follow the text cue on the screen to perform
the remaining four ventilations. When the
calibration is complete Omni will display
DONE.
7. Press SAVE.
RESETTING OMNI’S PRESSURE SENSOR
It is recommended that the Omni’s pressure
sensor is reset at altitudes greater than 1000
ft. to avoid inaccurate compressions and
ventilations readings. Perform the sensor reset
procedure only as part of the initial calibration
process.
1. Select RESET within the compression or
ventilation calibration menu.
After resetting the sensor, Omni will display
DONE.
2. Select SAVE to inish.
BP ARM
Use the BP Arm menu to gain access to
additional current settings, feedback graphics
and navigation menus that control all the
features of the blood pressure arm.
CURRENT SETTINGS
Current settings information is listed in the upper
center of the screen.
FEEDBACK GRAPHICS
The feedback graph located in the lower center
of the screen displays, in real time, the BP cuff
pressure.