Figure 32: Scheduler Page
Up to 8 different timers can be configured. For each timer, a
Start Date
and a
Start Time
must be
configured. The date and time will be referenced against the AccuLase-GPA™ real time clock, so
it is very important that the real time clock is correct. If it is not, be sure to set the clock correctly
by pressing the
Synchronize Time
button on the toolbar before setting timers. The date is in the
format YYYY-MM-DD. The time is in 24-hour format. The frequency at which the scheduled
analysis will reoccur, in hours and minutes, can be set in the fields below the start date and time.
If the frequency is set to 0 hours and 0 minutes, the scheduled analysis will not run as per the
schedule. A frequency MUST be set in order for the scheduled run to run as desired. The run that
is to be triggered by the timer can be chosen by placing a checkmark in the checkbox of the desired
run in the
Trigger Analysis
section of the timer configuration.
If the AccuLase-GPA™ is currently in the middle of running an automatic sequence at
the precise time the chosen analysis is scheduled to be run, the automatic sequence will
first complete, and THEN the scheduled run will be carried out. As a result, scheduled
analyses may not begin at precisely the time they are scheduled to be initiated.
4.4.9 Alarm Setup Page
The Alarms Setup page is used to configure hardware and calibration related alarms. It is shown
in Figure 41.