41
on the Calibration page are read-only. Additionally, calibration runs cannot be initiated from the
Calibration page – they must be initiated from the Calibration panel on the analyzer’s local display.
4.4.6 Run Setup Page
The Run Setup page is used to configure the various analysis runs that the AccuLase-GPA™ can
carry out. It is shown in Figure 38. There are a total of 8 different runs that can be configured on
the Run Setup page. When the page is first opened, the Run 1 setup is opened, and all seven
others are collapsed. To collapse a run, press the (-) button at the top right of its setup screen. To
expand a run, press the (+) on the right end of the Run title bar.
Figure 30: Run Setup Page
4.4.6.1 Run Definition
At the top of each run’s setup section are three fields that are used to define the parameters of
the run.
•
The
Run Name
is used to refer to the run everywhere throughout the AccuLase-GPA™
local display UI and in the web-based GUI. By default, Stream 1-4 are the run names used
for sample streams, Cal Span is used for the span calibration stream, and Cal Zero is used
for the zero calibration stream.
•
The
Enable
checkbox is used to enable the run and allow it to be carried out. If this
checkbox is not checked, the run will not run under any circumstances – it cannot be
initiated manually, cannot be set up as part of a stream sequence, and cannot be initiated
via a digital input or a timed frequency basis.
•
The
Type
drop-down menu is used to select the run type. Runs can either be one of four
types:
1.
Analysis
runs, where a concentration is calculated based on the analyzer’s
calibration parameters
2.
Zero Calibration
runs, which are used to establish the raw zero reading in the
analyzer’s calibration parameters