43
4.4.7 Sequencer Page
The Sequencer page is used to set up the AccuLase-GPA™ automatic sequence of analyses. It is
shown in Figure 39. The automatic sequence is a list of runs that the analyzer will perform
sequentially. When the defined automatic sequence is complete, the analyzer will return to the
beginning of the sequence and carry it out again. An automatic sequence will not be interrupted
by a scheduled analysis, an analysis initiated remotely via Modbus or a digital input, or an analysis
initiated manually via the Run Setup page – these runs will only be carried out once the currently
in progress sequence is complete. Further, once any run triggered by schedule, remotely, or
manually is completed, the analyzer will automatically return to the automatic sequence.
Figure 31: Sequencer Page
To add a run to the automatic sequencer, simply drag the desired run from the
Available Points
list and drop it in the
Automatic Sequencer Items
list. The same run can be placed in the automatic
sequencer list more than once if desired. Only runs that have been enabled in the Run Setup page
will run as part of the automatic sequence. To remove a run from the automatic sequencer, simply
click on it to highlight it, then hit the Delete key on the keyboard to remove it.
4.4.8 Scheduler Page
The Scheduler page is used to schedule runs to be initiated at a specific time. See Figure 40.