5-10
Chapter 5 Initial Settings
Measurement Related Setup
3
The measurement source of BP/NIBP mean value can be selected from waveform or calculation.
[Calc.]: Calculates the mean BP from the following calculation. Mean BP = (Systolic BP + Diastolic BP x 2) /3
[Wave]: In case of NIBP, maximum amplitude of oscillograph will be the mean value (MAP). In case of BP,
the following measurement will be performed.
NOTE
The NIBP specification stated on "Chapter 14 Specification" is based on the validation
when [Wave] is selected.
4
Set the "Arrhythmia Analysis Filter".
[Disp. Waveform]: The filter selected on "Admit/Discharge" menu or "ECG" setup menu will be set.
[Fixed]: The filter will be fixed to 0.5 Hz to 40 Hz.
NOTE
When [Disp. Waveform] is selected, the filter will be set according to the selection on
[Menu > Parameter > ECG]. If [Diag.] is selected, the filter will be 0.5 Hz to 40 Hz which
is the same with [Fixed].
5
Set the "Synchronized Mark/Tone".When [Auto] is selected for "Synchronized Mark/Tone", the priority of the
synchronizing parameter can be set.
[ECG]: The synchronizing priority will be set in the order of ECG>SpO
2
>BP. The synchronized tone will set
to [ON].
[SpO
2
]: The synchronizing priority will be set in the order of SpO
2
>ECG>BP. The synchronized tone will be
set to [ON].
6
Set the display priority of the parameter to be displayed inside the HR/PR numeric data box.
This priority setting will be applied when [Auto] is selected for "HR/PR", or when [HR/PR] user key is used to
switch the HR/PR source.
Select the priority order from the dropdown list.
For example, if [ECG/SpO
2
/BP] is selected, HR/PR source will be set in the priority of ECG>SpO
2
>BP.
7
Set the "Catheter Manufacturer for CC Input".Press one of the keys, and enter the manufacturer name on the
displayed window. (Max. 8 characters)
Example of BP Measurement
1: Systolic BP
2: HR interval
3: Mean BP (MEAN)
4: Diastolic BP
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......