9-6
Chapter 9 Maintenance Check
Daily and Periodic Check
No
Check Items
Check Procedure
Criteria
6. Respiration (Impedance Measurement)
01
Respiration Rate
display accuracy
Input the respiration signal of 20 Bpm with the
following setting from the simulator, and check
the displayed RR value on the monitor.
Baseline Impedance : 1500
Ω
,
Detection Lead: II (LL), Respiration Waveform:
Normal, Amplitude: 1
Ω
The displayed RR should be within 20 Bpm ±
3 Bpm.
No
Check Items
Check Procedure
Criteria
7. Invasive Blood Pressure
01
BP Zero
Input the blood pressure of 0 mmHg from the
simulator, perform zero balance, and check the
BP value on the monitor.
Zero balance should be properly performed,
and the displayed BP value should be within
0 mmHg ± 1 mmHg.
02
BP value display
accuracy
Set the BP value to 250 mmHg on the simulator,
and check the displayed BP value on the
monitor.
The displayed BP value should be within 250
mmHg ± 5 mmHg.
No
Check Items
Check Procedure
Criteria
8. SpO
2
01
Oxygen Saturation
Display Accuracy
Input the oxygen saturation signal of 90%SpO
2
from the simulator, and check the displayed
oxygen saturation value on the monitor.
The displayed oxygen saturation value
should be within 90%SpO
2
± 2%.
02
Pulse Rate Display
Accuracy
Input the pulse rate signal of 60 bpm from the
simulator, and check the displayed HR value on
the monitor.
The displayed heart rate should be within 60
bpm ± 3 bpm.
No
Check Items
Check Procedure
Criteria
9. NIBP
01
NIBP Test
Connect the 500 ml tank, and perform the NIBP
test under the test menu.
All the test results should be "OK".
02
BP Measurement
Error
Set the simulator to 120 mmHg for SYS, 80
mmHg for DIA, 90 mmHg for MAP, and perform
the NIBP measurement.
The displayed BP value should be within 120
mmHg ± 10 mmHg for SYS, 80 mmHg ±10
mmHg for DIA, 90 mmHg ±10 mmHg for
MAP.
03
Pulse Rate
Measurement Error
Set the pulse rate to 60 bpm on the simulator,
and perform the NIBP measurement.
The displayed pulse rate should be within 60
bpm ± 3 bpm.
No
Check Items
Check Procedure
Criteria
10. Temperature
01
Temperature display
error
Input the temperature signal of 37°C from the
simulator, and check the displayed temperature
value on the monitor.
The displayed temperature value should be
within 37°C ± 0.2°C.
No
Check Items
Check Procedure
Criteria
11. Cardiac Output (Blood Temperature, Injectate Temperature)
01
Blood Temperature
Measurement
Input the blood temperature signal of 37°C from
the simulator, and check the displayed blood
temperature value on the monitor.
The displayed blood temperature value
should be within 37°C ± 0.3°C.
02
Injectate Temperature
Measurement
Input the injectate temperature signal of 15°C
from the simulator, and check the displayed
injectate temperature value on the monitor.
The displayed injectate temperature value
should be within 15°C ± 0.5°C.
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......