Chapter 4 Connection to the External Devices
Ventilator Alarm Input
airway pressure upper limit alarm, high continuous pressure alarm, O
2
concentration
lower limit alarm, expiratory minute volume upper/lower limit alarm, apnea alarm, gas
supply alarm, air supply alarm, O
2
supply alarm, battery alarm, limited battery alarm, no
battery alarm, overrange alarm
SERVO-i:
airway pressure upper limit alarm, high continuous pressure alarm, O
2
concentration
lower limit alarm, expiratory minute volume upper/lower limit alarm, apnea alarm, gas
supply alarm, O
2
supply alarm, battery alarm, no battery alarm, limited battery alarm,
overrange alarm, expiratory cassette disconnected alarm, backup ventilation alarm,
regulation pressure limited alarm, respiratory rate alarm, PEEP low alarm, EtCO
2
upper
limit alarm, EtCO
2
lower limit alarm
SERVO-s:
airway pressure upper limit alarm, high continuous pressure alarm, O
2
concentration
lower limit alarm, expiratory minute volume upper/lower limit alarm, apnea alarm, gas
supply alarm, air supply alarm, O
2
supply alarm, backup ventilation alarm, respiratory
rate alarm, PEEP low alarm
PB 740/PB 760/PB 840:
The PB740/PB760/PB840 acquires alarm information from the nurse call port. The
ventilator alarm that cannot be acquired from the nurse call port is not guaranteed. For
corresponding alarm, refer to the service representative of the ventilator manufacturer.
This equipment is not compatible to the following alarms generated on the Evita 4 / Evita XL
/ Evita 2 dura.
O
2
monitoring disabled alarm, CO
2
alarm disabled alarm, Oximeter alarm disabled
alarm, Neo. volume measurement inoperable alarm, Minute volume alarm disabled
alarm, Minute volume alarm low off alarm, Tidal volume alarm high off alarm, Apnea
alarm off alarm, Nebulizer active alarm
When the Evita 2 dura/Evita 4/Evita XL is connected, there is a communication delay of 3
seconds between the DS-8007 System and the Evita ventilator.Therefore, if the alarm
generated at the ventilator is resolved within 3 seconds, the ventilator alarm may not be
generated at the DS-8007 System.
When connecting the VELIA, ASTRAL, VS ULTRA, refer to the respective operation
manual, and check which ventilator alarm will be output.
When the VELIA, ASTRAL, VS ULTRA ventilator is connected, and the ventilator power is
turned OFF, alarm will not generate on the DS-8007 System. If the connection cable is
disconnected from the ventilator, <Vent. Alarm> will generate, but <VENT_COMM> alarm
will not generate on the DS-8007 System.
CAUTION
The ventilator operation should be performed by well-trained and authorized personnel.
Only one ventilator can be connected to each DS-8007 system. Do not connect more than
one ventilators.
When connecting DS-8007 System and a ventilator, use only the specified connection
cable.
Make sure that the ventilator is connected to the specified connector on the DS-8007
System.
When connecting the cable, make sure that the main power of this system and the ventilator
is OFF.
On the DS-8007 system, the Status II connector and serial connector (COM) settings are
saved on the AC Unit (DSA-81).
If the DSA-81 is disconnected from the DS-8007 during ventilator monitoring, the ventilator
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......