8-3
Chapter 8 Cleaning/Disinfecting/Storing
Cleaning the Equipment and Sensors
2
Wipe the touch panel using a cleaning cloth.
3
Press again the [Key Lock] key for more than 2 seconds.
The message will disappear, and the key locked condition will be canceled.
Housing
Cleaning
Wipe using a tightly squeezed cloth saturated with diluted neutral detergent. Then wipe with a dry cloth.
Disinfection
Wipe with a cloth dampened with one of the following chemicals. Then, wipe off with dry cloth.
CAUTION
Clean the equipment frequently so stains can be removed easily.
To prevent injury, it is recommended to wear gloves when cleaning the equipment.
When cleaning or disinfecting, do not allow chemical solution to enter the equipment or
connectors.
Do not use organic solvents, thinner, toluene or benzene to avoid damaging the resin case.
Do not polish the equipment with abrasive, chemical cleaner, alkaline or acidic detergent.
Otherwise, the surface resin or paint coating may be damaged, resulting in discoloration,
scratches, and other problems.
For the precautions for storing and handling the chemicals, refer to the instruction manual
for the respective chemical.
NIBP Cuff and Air Hose
To clean the cloth bag of the cuff, remove the rubber bag and wash it using neutral detergent.Make sure that the cuff
is dry before placing back the rubber bag inside.
For procedure to clean/disinfect the reusable cuff, refer to the instruction manual for the respective cuff.
Do not reuse/resterilize the disposable cuff.
Usable Cloth:
*Soft cloth (cotton)
*Soft non-woven cloth (pulp, rayon, polyethylene, etc.)
Chemicals:
*Glutaral 2%
*Alcohol (ethanol, isopropyl alcohol for disinfection)
*Benzalkonium Chloride 0.2%
*Benzethonium Chloride 0.2%
*Alkyldiaminoethylglycine Hydrochloride 0.5%
Usable Cloth:
*Soft cloth (cotton)
*Soft non-woven cloth (pulp, rayon, polyethylene, etc.)
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......