9-9
Chapter 9 Maintenance Check
CO2 Calibration (HCP-810/HCP-820)
CO
2
Calibration (HCP-810/HCP-820)
This section describes about the procedure of CO
2
gas calibration.
Perform calibration for the following case.
When the measurement time exceeds 1,200 hours from the first use.
When 1 year has elapsed from the last calibration; or the accumulated EtCO
2
measurement time exceeds 4,000
hours, whichever comes earlier.
When error occurs to the measurement reading.
1
Press the [Menu], [CO
2
] ("Parameter"), [CO
2
Calibration] to display the CO
2
calibration screen.
2
Press the [Start Cal] key and conduct calibration
according to the displayed messages.
3
The message, <Feed CAL. GAS> will be displayed.
Press the injection button and inject the calibration gas.
4
The message, <Cal. Gas can be removed> will be
displayed. Stop pressing the injection button to cease
the injection.
5
The message, "CAL. OK" will be displayed. "Last Cal. Date" will be updated to the current date.
If any of the following messages is displayed, start the procedure again from step 2. <CAL. error>, <CAL
GAS error>, <Auto Zero fail>, <No stable gas flow>, <CAL. failure>
6
Press the [Cal Complete] key to end the calibration.
CAUTION
Perform the calibration 5 minutes after turning ON the power on the HCP-810/HCP-820.
Do not disconnect the sampling tube during calibration.If the sampling tube is
disconnected, calibration will cease.
Conduct CO
2
calibration for the following case.
When the accumulated measurement time exceeds 1,200 hours from the first use.
However, if the first calibration was performed before the accumulated measurement
time reaches 720 hours, another calibration is required when the accumulated
measurement time exceeds 1,200 hours from the first calibration.
When 12 months has elapsed or the accumulated measurement time has exceeded
4,000 hours from the previous calibration.
When EtCO
2
measurement is not stable or accuracy is degraded compared with other
measuring device.
When the patient monitor was not used for a while, or when EtCO
2
was not measured
for a while.
When a message, "Calibrate the CO
2
unit (HCP-810/HCP-820)" or "The periodic
calibration of the CO
2
unit (HCP-810/HCP-820) is approaching" is displayed at power
ON.
Dispose of calibration gas according to the regulation of each medical institution.
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......