5-7
Chapter 5 Initial Settings
Alarm Related Setup
(
Operation Manual "Vital Alarm Message" P11-1)
(
Operation Manual "Vital Alarm Message (DS-LAN Standard Setup)" P11-3)
2
Press the [Setup] key to display the alarm level setup window.
(shown on right)
3
Press one of the [Numeric Data], [Arrhythmia], [Technical] key.
The window will change according to the selected alarm
group.
4
Press the
/
keys.
The page will switch.
5
Select the alarm level from [S]/ [H]/ [M]/ [L]/ [N] for each parameter.
NOTE
Only the displayed alarm level can be selected.
Press the [Initialize] key to initialize the alarm level setting.
13
Set the operation for the alarm indicator located at the upper part of the main unit.
NOTE
The flashing colors of the alarm indicator are fixed for each alarm level and cannot be
changed.
Ventilator Alarm, Level H: Red
Level M: Yellow
Level L: Blue
1
Select [ON]/ [OFF] for each level.
When not using the alarm indicator function, select [OFF].
2
Press the [Indicator Test] key to test the flash pattern.
3
Select from [All ON] or [All OFF].
[All OFF]: Alarm indicator function will be turned OFF for all levels.
[All ON]: Alarm indicator function will be turned ON for all levels.
4
Set the "Synchronize with HR/RR".
[Sync. to HR]: The alarm indicator will flash in green synchronizing to HR.
[Sync. to RR]: The alarm indicator will flash in green synchronizing to RR.
[OFF]: The alarm indicator will not synchronize with HR/RR.
NOTE
If asystole alarm generates while [Sync. to HR] is selected, and [OFF] is selected for
alarm indicator function, the green LED of alarm indicator will remain lit. When PR
synchronized mark is displayed, the LED on the alarm indicator will not flash.
When [Sync. to RR] is selected and RR synchronized mark other than impedance is
displayed, the alarm indicator will not flash.
To turn OFF the alarm indicator operation all at once, press the [All OFF] key.
When the alarm is generated, and the alarm indicator is lit, the indicator will not flash
synchronizing to the HR/RR even if "Synchronize to HR/RR" is set.
Summary of Contents for Dynascope 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...Contents ...
Page 14: ...Contents ...
Page 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Page 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Page 52: ...2 2 Chapter 2 Network System Construction Contents ...
Page 62: ...3 2 Chapter 3 Using the External Media Contents ...
Page 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Page 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Page 86: ...5 2 Chapter 5 Initial Settings Contents ...
Page 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Page 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Page 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Page 153: ......
Page 154: ......