FLIR Detection
4. FLIR identiFINDER R300 Modes
Identi cation Con dence
ation Con dence
ation Con dence
Nuclide Names
Names
Names
Nuclide Usage
Nuclide Severity
Reference Number of
the Saved Record
Figure 96.
Identi ication Mode: Identi ied radionuclides
Continue
Acquire more data to improve the identi ication. The acquisition continues for the duration
you speci ied (see
6.4
, p.
116
) no matter whether you limited the duration or the minimum
number of counts for the initial acquisition.
Spectrum
Inspect the acquired spectrum.
This command is available only after you entered the password or if the password pro-
tection is off.
Please refer to
6.2
, p.
106
for options and commands for spectrum viewing. After you return
from the spectrum, you can
Continue
to acquire data, or
Exit
the identi ication mode.
4.4
Automated Mode
In automated mode, the FLIR identiFINDER R300 collects data for a radionuclide identi ication at
speci ied time intervals or in response to a dose rate warning. You can switch this mode on or off
and specify the interval (see
6.20
, p.
135
).
For each measurement, data will be collected until an instrument-speci ic number of gamma counts
has been registered (cf. “Dynamic” identi ication duration, see
6.4
, p.
116
).
In low-radiation environments, this can take quite a while, even longer than the interval you
speci ied. The next scheduled measurement will be omitted in such cases.
The start of an automated measurement will be announced 60 s in advance by a countdown shown
instead of the current time on the display of the FLIR identiFINDER R300 (Figure
97
, p.
94
).
identiFINDER
®
R300/en/2014.4(13623)/Feb2015
93