![FLIR identiFINDER R300 User Manual Download Page 117](http://html1.mh-extra.com/html/flir/identifinder-r300/identifinder-r300_user-manual_2306030117.webp)
FLIR Detection
6. FLIR identiFINDER R300 Menu Reference
This screen (Figure
123
, p.
117
) groups options that in luence the identi ication of nuclides (see
4.3
, p.
90
).
Figure 123.
Upper items of the identi ication settings
Settings and Commands
Duration
Specify the duration of data acquisition for the nuclide identi ication. Allot more time to the
identi ication measurement if:
• you want to identify a weak source.
• you are at a place with high background radiation.
• you want to identify a mixture of radionuclides.
• you want to identify a shielded source.
You can specify the durations “Dynamic” or from 1 s to 999 s (16 m 39 s).
If you specify a duration of “Dynamic”, the FLIR identiFINDER R300 will collect data un-
til 4800 gamma counts have been registered. This may take a while especially if the
radiation level is near the background level.
Extra Duration
Specify the duration of continued data acquisition for the nuclide identi ication. This dura-
tion is applied when you use the
Continue
command in identi ication mode (see
4.3
, p.
90
),
no matter whether you speci ied the duration or the minimum number of counts above.
You can specify an extra duration from 1 s to 999 s (16 m 39 s).
Con dence
Specify if you want the level of con idence of a nuclide identi ication shown or hidden in iden-
ti ication results. The con idence is expressed on a scale from 0 (poor) to 10 (excellent).