FLIR Detection
8. Using the FLIR identiFINDER R300 Web Server
Start
Starts the acquisition of data.
The FLIR identiFINDER R300 interrupts other operations during the acquisition. You
will be noti ied on the instrument’s screen (see
, p.
The spectrum chart and the information shown below the chart are updated every second
(Figure
219
, p.
213
).
The current spectrum is not cleared before recording data, that is, new measurements
are added to the existing spectrum. To record a pristine spectrum, use the
Clear
com-
mand irst (see below).
The acquisition of spectrum data does
not
stop if you navigate to another web page or even
disconnect the FLIR identiFINDER R300 from the computer. The acquisition of spectrum data
stops in one of the following events:
• You end the acquisition with
Stop
.
• You end the acquisition on the instrument with
Stop
(see
, p.
).
We recommend to stop data acquisition before you exit the live spectrum to relieve the
FLIR identiFINDER R300 from processing spectrum data.
Stop
Stops the acquisition of data.
Clear
Clears the spectrum.
This command clears the current spectrum but does not stop the current acquisition of
data, if any.
Identify
Applies the identi ication algorithm to the current spectrum. The result is shown above the
chart (Figure
220
, p.
214
).
The identi ication result is stored in the FLIR identiFINDER R300’s database. The reference
numbers of identi ication results saved during this live spectrum session are shown in a sec-
tion below the chart.
Save
Saves the current spectrum into the database of the FLIR identiFINDER R300. The reference
numbers of spectra saved during this live spectrum session are shown in a section below the
chart.
Logarithmic
Switches to logarithmic scaling along the vertical axis of the spectrum chart.
identiFINDER
®
R300/en/2014.4(13623)/Feb2015
211