6 Mounting
PIOX S502ID
6.2 Transducers
23
UMPIOX_S502IDV1-2EN, 2020-01-24
6.2.2
Installation of the transducers
6.2.2.1
Orientation of the transducers and determination of the transducer distance
Observe the orientation of the transducers. If the transducers have been mounted properly, the engravings on them form
an arrow (see Fig. 6.6). The transducer cables show in opposite directions.
The transducer distance is measured between the inner edges of the transducers.
• Select the installation instructions of the supplied transducer mounting fixture.
6.2.2.2
Transducer arrangement
The transducers can be arranged in the mounting rails in different ways:
Fig. 6.6:
Orientation of the transducers and transducer distance
a – transducer distance
Fig. 6.7:
Transducer arrangement in mounting rails
a
reflection arrangement,
1 rail
reflection arrangement,
2 rails
diagonal arrangement,
2 displaced rails
diagonal arrangement,
2 parallel rails
Summary of Contents for PIOX S502ID
Page 1: ...PIOX S502ID UMPIOX_S502IDV1 2EN Operating instruction ...
Page 63: ......
Page 64: ......