7 Connection
PIOX S502ID
7.1 Transducers
35
UMPIOX_S502IDV1-2EN, 2020-01-24
7.1.5
Transducer module (SENSPROM)
The transducer module contains important transducer data for the operation of the transmitter with transducers.
If transducers are replaced, the sensor module has to be replaced as well.
• Disconnect the transmitter from the power supply.
• Insert the transducer module into the socket.
• Connect the transmitter to the power supply.
• Enter all parameters of the program branch
Parameter
.
• Start the measurement.
Notice!
The serial numbers of the transducer module and the transducer have to be identical. A wrong or incorrectly
connected transducer module will lead to incorrect measured values or to a measurement failure.
Fig. 7.8:
Connection of the transducer module
1 – transducer module
AV
AV
S
AGN
ARS
AR