3 General principles
PIOX S502ID
3.3 Acoustic penetration
11
UMPIOX_S502IDV1-2EN, 2020-01-24
3.3
Acoustic penetration
The pipe has to be acoustically penetrable at the measuring point. The acoustic penetration is given when pipe and fluid
do not attenuate the sound signal so strongly that it is completely absorbed before reaching the second transducer.
The attenuation caused by the pipe and the fluid depends on:
• kinematic viscosity of the fluid
• proportion of gas bubbles and solid particles in the fluid
• deposits on the inner pipe wall
• pipe material
The following requirements have to be met at the measuring point:
• pipe always completely filled
• no deposits of solid particles in the pipe
• no formation of gas bubbles
Observe the following notes on the selection of the measuring point:
Horizontal pipe
Select a measuring point where the transducers can be mounted laterally on the pipe, allowing the sound waves to
propagate horizontally in the pipe. Thus, solid particles on the bottom of the pipe or gas bubbles in the pipe's upper part
are prevented from influencing the propagation of the signal.
Vertical pipe
Select the measuring point at a pipe section where the fluid flows upward. The pipe has to be completely filled.
Notice!
Even bubble-free fluids can form gas bubbles when the fluid expands, e.g., before pumps and after great cross-
section extensions.
Fig. 3.4:
Recommended transducer mounting position
Fig. 3.5:
Disadvantageous transducer mounting position
Fig. 3.6:
Recommended transducer mounting position
Fig. 3.7:
Disadvantageous transducer mounting position