9 Measurement
9.1 Start-up mode
PIOX S502ID
2020-01-24, UMPIOX_S502IDV1-2EN
46
9
Measurement
9.1
Start-up mode
• Select the menu item
Setup + measure
.
• Press ENTER.
9.1.1
Input of pipe parameters
• Select the pipe from the scroll list.
• If the pipe is not in the scroll lost, select
Other pipe
.
• Press ENTER.
If the pipe is selected from the scroll list, the corresponding parameters for outer diameter, wall thickness and pipe
material are set automatically. If
Other pipe
is selected, the parameters have to be entered.
• Enter the outer pipe diameter.
• Press ENTER.
If the entered parameter is outside the range, an error message will be displayed. The limit is displayed.
Example: upper limit 1100 mm for the connected transducers and for a pipe wall thickness of 50 mm
Pipe wall thickness
• Enter the pipe wall thickness.
• Press ENTER.
Warning!
Installation, connection and start-up by unauthorized and unqualified personnel
This may result in personal or material damage or dangerous situations.
→ Any work on the transmitter has to be carried out by authorized and qualified personnel.
Caution!
Safety and accident prevention regulations for electrical systems and equipment
Failure to observe these regulations may lead to severe injuries.
→ Observe the safety and accident prevention regulations for electrical systems and equipment.
START\Setup + measure
START\Setup + measure\Select pipe
START\Setup + measure\Outer Diameter
START\Setup + measure\Wall Thickness
Outer Diameter
1100.0
MAXIMAL