6 Mounting
6.3 Temperature probe
PIOX S502ID
2020-01-24, UMPIOX_S502IDV1-2EN
26
6.3.2.2
Installation with FLEXIM clasp
• Cut the tension strap to length (pipe circumf at least 120 mm).
• Insert approx. 20 mm of the tension strap into the slot of the clasp.
• Bend the end of the tension strap.
• Position the temperature probe on the pipe (see Fig. 6.12).
• Place the tension strap around the temperature probe and the pipe.
• Push the tension strap through part (2) and (1) of the clasp.
• Tighten the tension strap and engage it in the inner hook of the clasp.
• Tighten the screw of the clasp.
Fig. 6.11: Clasp
Fig. 6.12: Temperature probe on the pipe
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
→ Debur sharp edges.
→ Wear the required personal protective equipment.
→ Observe the applicable rules.
Fig. 6.13: FLEXIM clasp