9 Measurement
PIOX S502ID
9.2 Measuring mode
49
UMPIOX_S502IDV1-2EN, 2020-01-24
9.2
Measuring mode
After the input and confirmation of the transducer distance, the transmitter changes automatically to the measuring mode.
If this display appears, a field calibration for the detected fluid has to be carried out.
9.2.1
Field calibration
The sound speed measurement has to be adjusted to the detected fluid. For this purpose, a field calibration has to be
carried out on the transmitter while the fluid to be detected is in the pipe.
• Press
until
CALIBRATE FLUID
is displayed.
• Press ENTER.
• Enter the defined PIN code.
• Press ENTER.
The measurement is stopped for 5 s during the calibration. The calibration result is displayed and the measurement
continues automatically.
Display during the field calibration
The current fluid parameters are displayed. The calibration was successful.
This display will be indicated if the measured sound speed considerably differs from the sound speed of the fluid to be
detected.
Possible reasons:
• During the field calibration, the fluid to be detected was not in the pipe.
• The measuring system was not correctly installed.
This display will be indicated if the sound speed cannot be measured.
Possible reasons:
• The pipe is not completely filled.
• The measuring system was not correctly installed.
?
NaOH
?
not calibrated
?
NaOH
?
->CALIBRATE FLUID
INPUT PIN-Code
CODE
0000
Calibrated with
1234 m/s
-230
Can not calibr.
out of range
Can not calibr.
no soundspeed