7 Connection
PIOX S502ID
7.4 Temperature probe
41
UMPIOX_S502IDV1-2EN, 2020-01-24
Connection of the temperature probe to the junction box
• Remove the blind plug for the connection of the temperature probe (see Fig. 7.16).
• Open the cable gland of the temperature probe. The compression part remains in the cap nut.
• Push the cable of the temperature probe through the cap nut and the compression part.
• Prepare the cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the reducer into the junction box.
• Screw the basic part into the reducer.
• Insert the cable into the junction box.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the temperature probe to the terminals of the junction box.
Fig. 7.16: Junction box
1 – connection of the extension cable
2 – connection of the temperature probe
Tab. 7.11: Terminal assignment (junction box)
terminal
extension cable (KL2)
temperature probe (KL1)
1
red
red
2
gray
red/blue
3
white
white
4
blue
white/blue
1
2
KL2
KL1