9 Measurement
9.2 Measuring mode
PIOX S502ID
2020-01-24, UMPIOX_S502IDV1-2EN
50
9.2.2
Display during the measurement
By pressing
the following displays will be indicated:
The current fluid temperature is displayed. Press
.
Important data of the ongoing measurement are displayed. The quality and precision of the measurement can be evaluated.
• Press
.
Tab. 9.1:
Status display and values of the status outputs
display
state
current output I1
binary output B1
binary output B2
fluid to be detected
present in the pipe
20 mA
closed
closed
other fluid present in the
pipe
4 mA
closed
open
no fluid present in the
pipe, but transmitter ready
for operation
4 mA
open
open
invalid temperature
measurement
4 mA
open
open
measurement interrupted
or in start-up mode
0 mA
open
open
no display
transmitter de-energized
0 mA
open
open
57%
– signal quality
40.0dB
– SNR
1550
– sound speed in m/s
?
NaOH
?
is in pipe
?
NaOH
?
not in pipe
?
NaOH
?
no Signal
?
NaOH
?
no temper. input
>START< specials
Setup + measure
?
NaOH
?
Ts=
24.64 C
?
NaOH
?
57%40.0dB1550