www.eu.fishersci.com/fisherbrand
36
water tank.
System automatically runs a rinsing process for a newly installed RO
membrane to clean up its preservative solution. You may not be able to
dispense water during rinsing. RO indicator light blinks, indicating water quality
has not reached preset quality standards. Once the new RO membrane is
thoroughly rinsed, water production will return to normal.
System goes to standby automatically once the water tank is full. System
displays "Tank Full". When water is being dispensed, system will automatically
restart production of water.
If an excessive amount of water is dispensed, water level inside the PE tank
may fall below the minimum set-point. Water dispensing is stopped
automatically. System will display "Tank Empty" on the display panel. Once the
water tank is refilled sufficiently, water can be dispensed again.
3.3.3 Flushing and registering U Pack Cartridge on the Dispenser
a) Water dispenser starts automatically after system is powered up.
b) Press Dispenser button on the dispenser to start degassing and purge the
water loop, press dispenser button again when the dispenser shows
resistivity is 18.2 M
Ω
.cm and no air bubbles come out of the dispenser.
c) Registering Cartridge: In ready mode, press
and
simultaneously to
enter user menu. Press
to move cursor to PACK and press
to reset
cartridge life. Press
and
simultaneously again to exit user menu.
PACK: 360 DAYS 0L
FILTER: 360 DAYS
UV: 720 DAYS 0Hr
FLOW: 1.50 L/min
UNIT: M
Ω
.cm
TIME: 2014-11-27 09:21:27
LANGUAGE/
语言:
ENGLISH
SERIAL NO: S4PPAL3701