3.3.6 Calibrating Dispense Rate
1. Use a 2 liter volumetric cylinder, dispense water for 60 seconds and get
accurate volume of the collected water.
2. Press
and
simultaneously to enter user menu.
3. Press
to move cursor to FLOW.
4. Enter water volume dispensed to calibrate: Press
to move cursor to
corresponding number and press
to increase, enter water volume
collected in liter.
5. Press
to return to user menu
6. Press
and
simultaneously to exit user menu
3.3.7 Setting up Clock on the Main System
System time is preset. If you need to change the clock of the system, please
follow the procedure below.
1) Press down START and RO buttons simultaneously to set up display
system time. Format is Year - Month - Day, and Hour : Minute : Second
2) Press CURSOR button to enter editing mode.
PACK: 360 DAYS 0L
FILTER: 360 DAYS
UV: 720DAYS 0Hr
FLOW: 1.50 L/min
UNIT: M
Ω
.cm
TIME: 2014-11-27 09:21:27
LANGUAGE/
语言:
ENGLISH
SERIAL NO: S4PPAL3701
Date & Time
2015-01-01
01 : 01 : 01