www.eu.fishersci.com/fisherbrand
8
1.4 Operation
Accu500 integrated water systems produce ultrapure water and RO water
directly from tap water. Most particles, ions and organic compounds are
removed through the RO membrane before going to the storage tank. When in
need of ultrapure (UP) water, RO water flows through H Pack for further
de-ionization, through a UV ultraviolet light chamber (optional) to kill bacteria
and destroy trace organic pollutants in water, through a polish cartridge to
remove the last trace of ions, then a 0.2
µ
m final filter to the outlet.
After powering-up, the system goes into operation mode when the START
button is pressed. If RO water produced does not meet the preset quality
requirements, the RO indicator light on the panel will blink. RO water is
discharged to the drain until it meets quality standard, then into storage in the
water tank.
To dispense RO water, press the RO button on the system to dispense RO
water, and press RO button again to stop dispensing.
To dispense ultrapure water, press the dispense button on the dispenser to
dispense ultrapure water, and press the dispense button again to stop
dispensing.