3.3.4 Installing and Registering the Final Filter
1) Open the attached 0.2 um final filter package.
2) Screw the final filter onto the remote dispenser till finger tight (no leaking at
dispensing). Do not over tighten it as it may damage threads.
3) Press the Dispenser button to flush out gas in the filter.
4) Press the Dispenser button again to stop dispensing.
If a final ultrafiltration filter for pyrogen removal is needed, please install the
device the same way as the final filter.
5) Registering Final Filters: In ready mode, press
and
simultaneously to
enter user menu. Press
to move cursor to FILTER and press
to
reset final filter life. Press
and
simultaneously again to exit user menu.
3.3.5 Registering the Dual Wavelength UV Lamp
In Ready mode, press
and
simultaneously to enter user menu, press
to move cursor to UV
and press
to reset UV lamp life. Press
and
simultaneously again to exit user menu.
PACK: 360 DAYS 0L
FILTER: 360 DAYS
UV: 720DAYS 0Hr
FLOW: 1.50 L/min
UNIT: M
Ω
.cm
TIME: 2014-11-27 09:21:27