4. Find the UV lamp chamber (see picture). Remove the UV lamp cover to
expose the UV lamp.
5. Unplug the UV lamp from its power cord. Carefully remove the old UV
lamp.
6. Wear gloves included in the new UV lamp package. Avoid direct skin
contact with the quartz glass of the UV lamp.
7. Carefully insert the new UV lamp into its chamber. When about 2/3 of the
lamp is inserted, hold the UV lamp and connect it to the ballast cable
connector (4-pin connector) as shown in the picture, and then gently
insert the UV lamp completely into the chamber.
8. Cap the UV chamber with the black mask (see picture).
9. Reinstall the system top cover and tighten the screws.
Register the UV lamp as described in 3.3.5
UV
Lamp
UV Lamp
Mask
Power
Cord
UV Lamp
Connector