9
98
Eversense CGM User Guide
Alert Descriptions and Actions
(continued)
Alerts
App Displa
y
Description
Vibration Motor
Appears every 60 minutes when the vibration
motor on your smart transmitter can no longer
provide on-body vibe alerts. You will continue to
get glucose readings up to 72 hours after receiving
the alert message. After 72 hours, you will receive
a Transmitter Error Alert every 20 minutes until
you replace the smart transmitter.
Low Sensor Temperature*
Appears every 20 minutes when the sensor
temperature is too low.
No glucose reading is displayed until the sensor
temperature is within normal operating conditions.
A
ctions
Contact Customer Support to have your smart
transmitter replaced immediately.
Go to a warmer environment to increase the sensor
temperature. Keep your smart transmitter turned on so
you will start receiving glucose values when the sensor
temperature is between 26 - 40 °C (81 - 104 °F).
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 98
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 98
2/26/20 12:59 PM
2/26/20 12:59 PM