Eversense CGM User Guide
Table 3 – Eversense Percentage of Readings in Agreement Overall in the PRECISION Study
Percent of CGM System Readings Within
CGM
System
Glucose
Range
(mg/dL)
Paired
CGM
and YSI
References
Percent
15/15%
of
Reference
Percent
20/20%
of
Reference
Percent
30/30%
of
Reference
Percent
40/40%
of
Reference
> 40/40%
of
Reference
Overall
15,170
85.4%
92.8%
98.1%
99.3%
0.7%
40 - 60
1,236
91.9%
96.0%
98.4%
99.3%
0.7%
61 - 80
2,003
87.3%
94.1%
99.1%
99.6%
0.4%
81 - 180
5,786
80.5%
89.9%
97.2%
99.0%
1.0%
181 - 300
3,566
84.8%
92.8%
98.1%
99.2%
0.8%
301 - 350
1,628
92.8%
97.5%
99.1%
99.9%
0.1%
351 - 400
951
91.5%
95.8%
98.6%
99.8%
0.2%
Performance was also measured by calculating the percentage of sensor glucose readings within 15 mg/dL or 15%
of the YSI reference. These tables show the percent agreement at multiple levels, at different glucose ranges, and
at different days during the sensor wear. Results in the glucose ranges of 80 mg/dL or less reflect the percentage of
values within mg/dL, and results in the glucose ranges over 80 mg/dL reflect the percentage within reference. As an
example, glucose values between 40 and 60 mg/dL were within 15 mg/dL of the reference value 91.9% of the time.