6
45
Eversense CGM User Guide
Calibration Phases
A. Initialization Phase (after 24-hour Warm-Up Phase)
During this phase, 4 fingerstick blood glucose meter tests are required.
•
The 4 calibration tests must be spaced 2 to 12 hours apart, and all 4 tests must be completed within a 36 hour
period. After 8 hours without a calibration entry, no glucose data will be displayed.
– 1
st
calibration = 24 hours after sensor insertion.
– 2
nd
calibration = 2 to 12 hours after 1st successful calibration.
– 3
rd
calibration = 2 to 12 hours after 2nd successful calibration.
– 4
th
calibration = 2 to 12 hours after 3rd successful calibration.
•
Glucose readings will start displaying in the app a few minutes after the 2nd calibration is successfully completed.
IMPORTANT:
If your smart transmitter is not turned on and paired with the Eversense App and sensor, the
system is not able to prompt you to calibrate.
Re-Entering Initialization Phase
The following will cause the system to re-enter Initialization Phase.
•
Not completing a calibration test within a 12-hour period during the Initialization Phase.
•
Not completing all 4 calibration tests within 36 hours during the Initialization Phase.
•
Not completing 2 calibration tests within a 24-hour period during the Daily Calibration Phase
(see
B. Daily Calibration Phase
).
•
When the last several blood glucose meter measurements are significantly different than the sensor glucose values.
•
If the smart transmitter is out of battery power for more than 16 hours.
•
When you receive a Sensor Check Alert.
•
Six hours after you receive a Sensor Suspend Alert.
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 45
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 45
2/26/20 12:58 PM
2/26/20 12:58 PM